• Title of article

    Spectroscopic hemoglobin and bilirubin measurement on optically opaque particulate fluids

  • Author/Authors

    Brandl، نويسنده , , Martin and Kellner، نويسنده , , Karlheinz and Posnicek، نويسنده , , Thomas and Bado، نويسنده , , Itka and Falkenhagen، نويسنده , , Dieter، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    7
  • From page
    711
  • To page
    717
  • Abstract
    Extracorporeal blood purification systems based on specific micro-adsorbents can be used for selective elimination of toxic substances from the patient. A suspension of blood plasma with micro-adsorbents is recirculated in the extracorporeal detoxification circuit. A safety system detecting erythrocyte damage by hemoglobin release as well as measuring the bilirubin elimination rate in the adsorbent circuit was developed. The sensor principle is based on broadband illumination and reflectivity measurement in the tubing where the adsorbent–plasma suspension is recirculated. From the specific reflection spectrum the hemoglobin and bilirubin content is calculated. Further, the effects of cross-sensitivity of hemoglobin and bilirubin on the detection accuracy are studied.
  • Keywords
    Blood purification , Hemoglobin sensor , Bilirubin sensor , optical sensors
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2013
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1442109