• Title of article

    Performance of potentiometric oxygen sensors with LSM-GDC composite electrode in liquid LBE at low temperatures

  • Author/Authors

    Lim، نويسنده , , J. and Manfredi، نويسنده , , G. and Mariën، نويسنده , , A. and Van den Bosch، نويسنده , , J.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    7
  • From page
    1048
  • To page
    1054
  • Abstract
    Potentiometric oxygen sensors integrated with LSM (strontium-doped lanthanum manganite)-GDC (gadolinium-doped ceria) composite electrode have been developed for use in liquid LBE at low temperatures. The performance of these sensors has been evaluated in oxygen saturated LBE in the temperature range of 200–400 °C. Their performance was compared with that for platinum electrode and Bi/Bi2O3 electrode. The sensor with LSM-GDC composite electrode performed well down to 200 °C by using a voltmeter with high input resistance (>10 GΩ), while those with platinum electrode and Bi/Bi2O3 electrode showed significant deviations at 200 °C.
  • Keywords
    MYRRHA , Lead-bismuth eutectic , Potentiometric oxygen sensor , LSM-GDC composite electrode
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2013
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number

    1443089