• Title of article

    Preparation and bioelectrochemical responses of the poly(m-phenylenediamine) glucose oxidase electrode

  • Author/Authors

    Zhou، نويسنده , , Haihui and Chen، نويسنده , , Hong and Luo، نويسنده , , Shenglian and Chen، نويسنده , , Jinhua and Wei، نويسنده , , Wanzhi and Kuang، نويسنده , , Yafei، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    7
  • From page
    224
  • To page
    230
  • Abstract
    A new technique termed the “pulse galvanostatic method (PGM)” is described for constructing a glucose sensor by electropolymerization of m-phenylenediamine in the phosphate buffer solution containing m-phenylenediamine and glucose oxidase (GOx). The conditions for fabrication of the poly(m-phenylenediamine) glucose oxidase (PMPD GOx) electrode were optimized as follows: ton/toff, 1; frequency, 10; mean current density, 1 mA cm−2; concentration of the GOx, 5 g l−1. The morphology of PMPD GOx electrode prepared by PGM under the optimized conditions shows porous structure. The PMPD GOx electrode has short response time, high response current, good selectivity and stability. The maximum response current density and Michaelis–Menten constant of the PMPD GOx electrode, based on the calculation of the kinetic parameters, are 231.4 μA cm−2 and 15.33 mM, respectively.
  • Keywords
    Pulse galvanostatic method , Poly(m-phenylenediamine) , Enzyme electrode , Glucose oxidase
  • Journal title
    Sensors and Actuators B: Chemical
  • Serial Year
    2004
  • Journal title
    Sensors and Actuators B: Chemical
  • Record number