• Title of article

    Sequential-injection stripping analysis of nifuroxime using DNA-modified glassy carbon electrodes

  • Author/Authors

    Diab، نويسنده , , Nizam and AbuZuhri، نويسنده , , Ali and Schuhmann، نويسنده , , Wolfgang، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    7
  • From page
    57
  • To page
    63
  • Abstract
    The voltammetric behavior of nifuroxime was investigated comparing stationary voltammetric methods with the recently proposed sequential-injection stripping analysis (SISA), by using cyclic voltammetry (CV) and differential-pulse voltammetry at bare and DNA-modified glassy carbon (GC) electrodes. In cyclic voltammetry, reduction of nifuroxime at DNA-modified electrodes gives rise to a well-defined peak, and in contrast to bare GC surfaces, a re-oxidation peak could be observed. Optimization of the pre-concentration process at the DNA-modified surface led to a significant enhancement of the voltammetric current response, a better defined peak shape and an improved dynamic range. Based on this optimized voltammetric procedure, SISA has been evaluated for the determination of nifuroxime. The flow-system significantly facilitates the regeneration of the DNA-modified electrode surface, hence diminishing problems related to accumulation and memory effects. The linear detection range could be extended to 65 μM with a detection limit (3 s) of 0.68 μM, which corresponds to an absolute amount of 21 ng nifuroxime.
  • Keywords
    Nifuroxime , Sequential-injection stripping analysis , DNA-modified electrode , Modified electrode
  • Journal title
    Bioelectrochemistry
  • Serial Year
    2003
  • Journal title
    Bioelectrochemistry
  • Record number

    1450696