• Title of article

    Determination of plasma testosterone using a simple liquid chromatographic method

  • Author/Authors

    Ng، نويسنده , , Bee Hong and Yuen، نويسنده , , Kah Hay، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    6
  • From page
    421
  • To page
    426
  • Abstract
    A simple and sensitive high-performance liquid chromatographic (HPLC) method using ultraviolet detection was developed for the determination of testosterone in human plasma. Testosterone and the internal standard, griseofulvin, were extracted from 0.50 ml plasma sample using a mixture of dichloromethane–2,2,4-trimethylpentane (3:2, v/v). The mobile phase, consisted of 0.02 M sodium dihydrogenphosphate–acetonitrile–methanol (51:47:2, v/v) adjusted to pH 3.1 and delivered to a C18 analytical column (150×4.6 mm I.D., 4 μm particles) at a flow-rate of 1 ml/min while the detection wavelength was set at 240 nm with a sensitivity range of 0.005 a.u.f.s. The method has a quantification limit of 1.6 ng/ml. Recoveries of testosterone were all greater than 92% over the linear concentration range of 1.6–400 ng/ml while that of griseofulvin was approximately 95%. The within- and between-day RSD values were all less than 8% while the accuracy values ranged from 96.0 to 106.0% over the concentration range studied. The method was applied to the analysis of early morning plasma testosterone levels of 12 healthy human male volunteers. The levels were found to range from 3.1 to 8.4 ng/ml, within the normal range reported in the literature.
  • Keywords
    Testosterone
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2003
  • Journal title
    Journal of Chromatography B
  • Record number

    1455730