• Title of article

    Validation of a simple liquid chromatography assay for creatine suitable for pharmacokinetic applications, determination of plasma protein binding and verification of percent labeled claim of various creatine products

  • Author/Authors

    Persky، نويسنده , , Adam M. and Hochhaus، نويسنده , , Günther and Brazeau، نويسنده , , Gayle A.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    9
  • From page
    157
  • To page
    165
  • Abstract
    Creatine has been quantified in various tissues by a range of methodologies. This paper reports on the development and validation of a simplified HPLC assay to determine plasma creatine, plasma protein binding of creatine, creatine in microdialysate and creatine in over-the-counter products. An isocratic, reversed-phase (C18) HPLC assay, using potassium phosphate monobasic (pH 4) as a mobile phase, was validated in human plasma and microdialysis perfusion fluid (normal saline). The lower limit of quantification for the assay was 1 mg l−1 in saline and 5 mg l−1 in plasma. The RSD was below 6% and accuracy was below 12% in both matrices. Protein binding in human plasma was found to be negligible (<10%). Over-the-counter creatine monohydrate products tested contained 100% creatine monohydrate. This assay was found to be suitable for pharmacokinetic studies and the assessment of plasma creatine and skeletal muscle microdialysate.
  • Keywords
    CREATINE
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2003
  • Journal title
    Journal of Chromatography B
  • Record number

    1455771