• Title of article

    Determination of 3-amino-5-mercapto-1,2,4-triazole in serum

  • Author/Authors

    Depree، نويسنده , , G.J. and Siegel، نويسنده , , P.D.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    4
  • From page
    359
  • To page
    362
  • Abstract
    A high performance liquid chromatography (HPLC) method using fluorescence detection to determine 3-amino-5-mercapto-1,2,4-triazole (AMT) levels in serum has been developed. Sample preparation involved treatment with tributylphosphine (TBP) to reduce disulfides formed during storage, precipitation of proteins with acetonitrile (ACN), and precolumn derivatization using the thiol reactive fluorescent probe monobromobimane (MBB). The conjugate (AMT-MBB) was resolved by gradient elution from a C18 reversed-phase column. The assay method was linear over a concentration range of 0.78–50 μg/ml and had a limit of detection (LOD) of 0.05 μg/ml AMT (10 μl injection). This method provides a sensitive and specific tool for the determination of AMT in serum and may have potential industrial hygiene application.
  • Keywords
    3-Amino-5-mercapto-1 , 2 , 4-Triazole
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2004
  • Journal title
    Journal of Chromatography B
  • Record number