• Title of article

    Determination of Tirofiban in human serum by liquid chromatography–tandem mass spectrometry

  • Author/Authors

    Oertel، نويسنده , , R and Kِhler، نويسنده , , A and Koster، نويسنده , , A and Kirch، نويسنده , , W، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    5
  • From page
    181
  • To page
    185
  • Abstract
    A liquid chromatography–tandem mass spectrometric (LC–MS–MS) method with a rapid and simple sample preparation was developed and validated for the determination of Tirofiban in biological fluids. Tirofiban in serum samples was extracted and cleaned up by using an automated solid phase extraction method. An external calibration was used. The mass spectrometer was operated in the multiple reaction monitoring mode (MRM). A good linear response over the range of 2–200 ng/ml was demonstrated. The accuracy for Tirofiban ranged from 94.8 to 110.8% within-day and from 103.0 to 104.7% between-day. The lower limit of quantification was 2 ng/ml. This method is suitable for pharmacokinetic studies.
  • Keywords
    Tirofiban
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2004
  • Journal title
    Journal of Chromatography B
  • Record number