• Title of article

    Sensitive assay for determining plasma tenofovir concentrations by LC/MS/MS

  • Author/Authors

    Delahunty، نويسنده , , Tom and Bushman، نويسنده , , Lane and Fletcher، نويسنده , , Courtney V.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    7
  • From page
    6
  • To page
    12
  • Abstract
    An LC/MS/MS assay for the determination of tenofovir (TNF) was developed and validated for use with the EDTA anticoagulated human plasma matrix. Heparin-treated plasma and serum matrices were also validated. After addition of adefovir as an internal standard, trifluoroacetic acid was used to produce a protein-free extract. Chromatographic separation was achieved with a Polar-RP Synergi, 2.0 mm × 150 mm, reversed-phase analytical column. The mobile phase was 3% acetonitrile/1% acetic acid, aq. Detection of TNF and the internal standard was achieved by ESI MS/MS in the positive ion mode using 288/176 and 274/162 transitions, respectively. The method was linear from 10 to 750 ng/ml with a minimum quantifiable limit of 10 ng/ml when 250 μl aliquots were analyzed. The usefulness of this LC/MS/MS method to routinely monitor plasma concentrations of TNF was demonstrated along with its ability to assist in the performance of pharmacokinetic studies.
  • Keywords
    Tenofovir (TNF) , Adefovir (PMEA) , LC/MS/MS , PLASMA
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2006
  • Journal title
    Journal of Chromatography B
  • Record number

    1462535