Title of article
Determination of mycophenolic acid glucuronide in microsomal incubations using high performance liquid chromatography–tandem mass spectrometry
Author/Authors
Mohamed، نويسنده , , Mohamed-Eslam F. and Harvey، نويسنده , , Stephen S. and Frye، نويسنده , , Reginald F.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
4
From page
251
To page
254
Abstract
A sensitive and specific HPLC–MS/MS method was developed for the analysis of mycophenolic acid glucuronide (MPAG) in incubations with human liver microsomes. Incubation samples were processed by protein precipitation with acetonitrile. MPAG and the internal standard phenolphthalein glucuronide were chromatographed on a C18 Synergi Fusion-RP column (100 mm × 2 mm, 4 μm) using gradient elution with a mixture of 1 mM acetic acid in deionized water and 1 mM acetic acid in acetonitrile at a flow rate of 0.22 mL/min. The mass spectrometer was operated with negative electrospray ionization and analysis was carried out in the single reaction monitoring (SRM) mode using the mass transitions of m/z 495 → 319 and m/z 493 → 175 for MPAG and phenolphthalein glucuronide, respectively. The MPAG calibration curve was linear over the concentration range of 1.0–20 μM. The within-day and between-day relative standard deviations ranged from 1.1 to 7.9% and accuracy was within 8%. The simple and reproducible method is suitable for measuring mycophenolic acid glucuronidation in microsomal incubations.
Keywords
Mycophenolic acid , Glucuronidation , drug metabolism
Journal title
Journal of Chromatography B
Serial Year
2008
Journal title
Journal of Chromatography B
Record number
1466144
Link To Document