• Title of article

    Total error profiling of a proinsulin time-resolved fluorescence immunoassay

  • Author/Authors

    De Pauw، نويسنده , , P.E. and Mackin، نويسنده , , R.B. and Goubert، نويسنده , , P. and Van Schravendijk، نويسنده , , C. and Gorus، نويسنده , , F.K.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    4
  • From page
    2403
  • To page
    2406
  • Abstract
    We applied total error profiling to evaluate the conversion of a known proinsulin (PI) enzyme-linked immunosorbent assay (ELISA) into a time-resolved fluorescence immunoassay (TRFIA). The formula and acceptance criteria proposed by the Ligand Binding Assay Bioanalytical Focus Group (LBABFG) of the American Association of Pharmaceutical Scientists (AAPS) were applied. We found that the expected dynamic range enlargement with TRFIA compared to ELISA ([0.5–240] versus resp. [0.7–98] pmol/L) is limited by an interference of C-peptide when present in the sample at high concentrations (>7000 pmol/L).
  • Keywords
    Total error profiling , time-resolved fluorescence , Proinsulin , C-peptide , diabetes , IMMUNOASSAYS , Validation
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2009
  • Journal title
    Journal of Chromatography B
  • Record number

    1467451