• Title of article

    Accuracy profile validation of a new analytical method for butane measurement using headspace-gas chromatography–mass spectrometry

  • Author/Authors

    Varlet، نويسنده , , V. and Smith، نويسنده , , F. and Augsburger، نويسنده , , M.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2013
  • Pages
    6
  • From page
    155
  • To page
    160
  • Abstract
    The aim of our study was to provide an innovative HS-GC/MS method applicable to the routine determination of butane concentration in forensic toxicology laboratories. The main drawback of the GC/MS methods discussed in literature concerning butane measurement was the absence of a specific butane internal standard necessary to perform quantification. Because no stable isotope of butane is commercially available, it is essential to develop a new approach by an in situ generation of standards. To avoid the manipulation of a stable isotope-labelled gas, we have chosen to generate in situ an internal labelled standard gas (C4H9D) following the basis of the stoichiometric formation of butane by the reaction of deuterated water (D2O) with Grignard reagent butylmagnesium chloride (C4H9MgCl). This method allows a precise measurement of butane concentration and therefore, a full validation by accuracy profile was presented.
  • Keywords
    Headspace gas extraction , In situ stable labelled isotope generation , Butane , Deuterated butane , Grignard reagent
  • Journal title
    Journal of Chromatography B
  • Serial Year
    2013
  • Journal title
    Journal of Chromatography B
  • Record number

    1470791