• Title of article

    High temperature and temperature programming in capillary electrochromatography

  • Author/Authors

    Djordjevic، نويسنده , , N.M and Fitzpatrick، نويسنده , , F and Houdiere، نويسنده , , F and Lerch، نويسنده , , G and Rozing، نويسنده , , G، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    8
  • From page
    245
  • To page
    252
  • Abstract
    In electrochromatography, solvent electrophoretic mobility and solute partitioning are temperature dependent processes. If temperature variations are controlled, solute selectivity and analysis times can be tailored. In this study the feasibility of temperature programming in capillary electrochromatography (CEC) was demonstrated using a reversed-phase CEC mode. The outcome of programmed separations was compared with isothermal, isocratic and isorheic (constant flow) separations. The combined effects of column temperature and mobile phase flow-rate changes during the separation run, resulted in up to a 50% reduction in the separation run time, without adversely affecting the quality of separation. For capillary electrochromatography, temperature programming may be a valuable alternative to solvent programming modes because of the great technical difficulties associated with carrying out solvent gradient elution.
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2000
  • Journal title
    Journal of Chromatography A
  • Record number

    1501962