• Title of article

    Ion chromatographic analysis of tetracyclines using polymeric column and acidic eluent

  • Author/Authors

    Ding، نويسنده , , Xiaojing and Mou، نويسنده , , Shifen، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    10
  • From page
    205
  • To page
    214
  • Abstract
    High-performance ion chromatography (HPIC) is first successfully used to analyze tetracycline antibiotics (TCs) in this work. The TCs are well separated on a solvent compatible polymeric cation-exchange column within 12 min. Isocratic elution with acetonitrile-hydrochloride is very advantageous for routine analysis. HPIC may be seen as a specific variant of the more common high-performance liquid chromatography (HPLC) for water-soluble and polar pharmaceuticals with low hydrophobicity. The detection limits (signal-to-noise ratio=3:1) of oxytetracycline (OTC), tetracycline (TC), chlortetracycline (CTC), doxycycline (DC) are 10, 10, 20 and 20 μg l−1, respectively. Samples are prepared by vortex mixing with an ethylenediaminetetraacetic acid disodium salt (Na2EDTA)–McIlvaine buffer (pH 4.0) solution and the mixture filtrates through a molecular weight cut-off filter. The method has been successfully applied to monitor the OTC removal rate through every reactor in the process of OTC manufacturing wastewater treatment by bio-chemical technology. It is also applicable to determine the TCs residues in milk and milk powder with satisfying results.
  • Keywords
    antibiotics , Tetracyclines
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2000
  • Journal title
    Journal of Chromatography A
  • Record number