Title of article
Effect of organic solvents in the on-line thermal lens spectrometric detection of chromium(III) and chromium(VI) after ion chromatographic separation
Author/Authors
?ikovec، نويسنده , , Mateja and Franko، نويسنده , , Mladen and Novi?، نويسنده , , Milko and Veber، نويسنده , , Marjan، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2001
Pages
7
From page
119
To page
125
Abstract
The effect of methanol, acetone and acetonitrile on the sensitivity, selectivity and the detection limits (LODs) of the determination of chromium species by ion chromatography was investigated. A collinear dual-beam thermal lens spectrometer was used for the direct detection of chromium complexes [pre-column derivatized Cr(III)–pyridine-2,6-dicarboxylic acid, and post-column derivatized Cr(VI)–1,5-diphenylcarbazide] following the ion chromatographic separation on a Dionex HPIC-CS5A solvent compatible column. Different amounts of organic solvents were added directly to the eluent (up to 30%) and to the post-column reagent (up to 60%) to improve the thermooptical properties of the solvents. Consequently, the sensitivity of the technique was increased by a factor of 2–3 and LODs of 0.1 and 10 μg dm−3 were achieved for Cr(VI) and Cr(III), respectively, when the eluent reaching the detector contained 30% of acetonitrile. The addition of organic solvents also resulted in significant changes in retention times, which improved the Cr(III)/Cr(VI) separation.
Keywords
Metal cations , Chromium
Journal title
Journal of Chromatography A
Serial Year
2001
Journal title
Journal of Chromatography A
Record number
1507964
Link To Document