Title of article
Gas chromatographic separation of fatty acid methyl esters on weakly polar capillary columns
Author/Authors
Yamamoto، نويسنده , , Kouhei and Kinoshita، نويسنده , , Akemi and Shibahara، نويسنده , , Akira، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2008
Pages
4
From page
132
To page
135
Abstract
It was found that weakly polar columns, routinely used in capillary GC for analyzing sterols, food additives, etc., can also be used for separating fatty acid methyl esters (FAMEs). On these columns, FAMEs elute in the order of their unsaturation. The equivalent chain-length value of methyl 22:6 is below 23.00. This means FAMEs within a carbon chain length, having up to six double bonds, elute before the next (one carbon longer) saturated FAME elutes. Peak identification is easy. Weakly polar columns are compatible in both GC and GC/MS systems.
Keywords
Separation of fatty acid methyl esters , Equivalent chain-length value , Capillary gas chromatography , Weakly polar capillary column , Fish oil capsule