• Title of article

    Determination of multiple pharmaceutical classes in surface and ground waters by liquid chromatography–ion trap–tandem mass spectrometry

  • Author/Authors

    Gruji?، نويسنده , , Svetlana and Vasiljevi?، نويسنده , , Tatjana and Lau?evi?، نويسنده , , Mila، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2009
  • Pages
    12
  • From page
    4989
  • To page
    5000
  • Abstract
    This paper describes development, optimization and application of analytical method for determination and reliable confirmation of nineteen pharmaceuticals from different therapeutic classes (antibiotics—β-lactams, cephalosporines, sulfonamides, macrolides and tetracyclines; benzodiazepines; antiepileptics and analgoantipyretics) in surface and ground waters at ng l−1 levels. Water samples were prepared using solid-phase extraction and extracts were analyzed by liquid chromatography–ion trap–tandem mass spectrometry with electrospray ionization in both positive and negative ionization mode. The efficiency of ten different SPE cartridges to extract diverse compounds from water was tested. The pH-value of the water sample, the volume of elution solvent and the sample volume were optimized. Matrix effect, especially pronounced for cephalexin and metamizole, was eliminated using matrix-matched standards. It was determined that extraction should be performed at pH ∼ 7.5, i.e. without pH adjustment, and at pH 3, depending on the analyte. Azithromycin, doxycycline and acetylsalicylic acid must be extracted in acidic environment, whereas extraction of paracetamol, ampicillin, erythromycin and metamizole should be performed without pH adjustment. Repeatability of the method was generally lower than 20%. The estimated limits of detection were in the range from 0.15 to 12.46 ng l−1. The method was applied to 26 water samples for monitoring of selected drug residues. Results revealed the presence of carbamazepine (80% of water samples), azithromycin (23%), as well as trimethoprim and paracetamol (both 15%). The most striking was the false positive signal of diclofenac in every analyzed water sample. Confirmation of the positive results was performed by repeated injection of the positive sample extracts using confirmatory method with additional transitions.
  • Keywords
    Multiclass pharmaceuticals , antibiotics , Liquid chromatography–tandem mass spectrometry , Ion trap , Solid-phase extraction , Water analysis
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2009
  • Journal title
    Journal of Chromatography A
  • Record number