    Sensitive high-performance anion-exchange chromatographic determination of paeoniflorin and albiflorin by pulsed amperometric detection after solid-phase extraction

  • Author/Authors

    Sim، نويسنده , , Hee-Jung and Jeong، نويسنده , , Ji-Seon and Kwon، نويسنده , , Ha-Jeong and Lee، نويسنده , , Yong-Moon and Hong، نويسنده , , Seon-Pyo، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    4
  • From page
    5302
  • To page
    5305
  • Abstract
    We developed a method to simultaneously determine paeoniflorin and albiflorin levels using high-performance anion-exchange liquid chromatography with pulsed amperometric detection (HPAEC-PAD). The main principle of our method includes solid-phase extraction step using Amberlite XAD-2 sorbent to remove sugars and to selectively determine glycosides by PAD. Under these conditions, the linear dynamic range was 0.01–100 μg/mL, and the albiflorin and paeoniflorin detection limits (S/N = 3) were 5 and 10 pg, respectively. The intra- and inter-day precisions (RSDs) were <5.07%, and the average recoveries from Paeoniae Radix and Si-ni-san ranged from 97.12 to 101.15%. Our method showed high selectivity, high sensitivity, and good repeatability for analyzing albiflorin and paeoniflorin in oriental medicinal preparation.
  • Keywords
    Anion-exchange chromatography , PAD , Paeoniae Radix , Paeoniflorin , Albiflorin , Solid-phase extraction
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2010
  • Journal title
    Journal of Chromatography A
  • Record number

    1513295