Title of article
A direct Capillary Liquid Chromatography with electrochemical detection method for determination of phenols in water samples
Author/Authors
C. and Segovia-Martيnez، نويسنده , , L. and Moliner-Martيnez، نويسنده , , Y. and Campيns-Falcَ، نويسنده , , P.، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2010
Pages
5
From page
7926
To page
7930
Abstract
A fast and direct method based on the use of Capillary Liquid Chromatography (LC) with electrochemical (EC) detection has been described for phenols pollutants in water samples. Concretely, phenol, o-cresol, 2-chlorophenol and bisphenol A have been selected as target analytes. The combination of Capillary LC with EC detection avoided the necessity of preconcentration steps typically used in environmental analysis. The sample injected volume was 2 μL. The achieved detection limits were between 1 and 2 μg/L and the linear dynamic range was up to 50 μg/L for all studied phenols. The precision and uncertainty were satisfactory. The analysis time per sample was 10 min. The proposed procedure has been proved useful for treated waters.
Keywords
Capillary LC , Electrochemical detection , Phenols , Treated water , Environmental field
Journal title
Journal of Chromatography A
Serial Year
2010
Journal title
Journal of Chromatography A
Record number
1513619
Link To Document