• Title of article

    A direct Capillary Liquid Chromatography with electrochemical detection method for determination of phenols in water samples

  • Author/Authors

    C. and Segovia-Martيnez، نويسنده , , L. and Moliner-Martيnez، نويسنده , , Y. and Campيns-Falcَ، نويسنده , , P.، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2010
  • Pages
    5
  • From page
    7926
  • To page
    7930
  • Abstract
    A fast and direct method based on the use of Capillary Liquid Chromatography (LC) with electrochemical (EC) detection has been described for phenols pollutants in water samples. Concretely, phenol, o-cresol, 2-chlorophenol and bisphenol A have been selected as target analytes. The combination of Capillary LC with EC detection avoided the necessity of preconcentration steps typically used in environmental analysis. The sample injected volume was 2 μL. The achieved detection limits were between 1 and 2 μg/L and the linear dynamic range was up to 50 μg/L for all studied phenols. The precision and uncertainty were satisfactory. The analysis time per sample was 10 min. The proposed procedure has been proved useful for treated waters.
  • Keywords
    Capillary LC , Electrochemical detection , Phenols , Treated water , Environmental field
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2010
  • Journal title
    Journal of Chromatography A
  • Record number

    1513619