• Title of article

    Microscale solid-phase extraction system for explosives

  • Author/Authors

    Smith، نويسنده , , Matthew and Collins، نويسنده , , Greg E and Wang، نويسنده , , Joseph، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    9
  • From page
    159
  • To page
    167
  • Abstract
    A simple, semi-automated, microcolumn solid-phase extraction (SPE) system is optimized for the extraction, preconcentration and HPLC analysis of seven different explosives and explosive derivatives contaminating seawater, river water and well water samples. The microcolumns were constructed from 1/16 in. O.D. PTFE tubing (1 in.=2.54 cm) packed with 0.5–1.5 mg of SPE material, LiChrolut EN or Porapak R. The extraction system consisted of two syringe pumps and several solenoid valves. Optimal detection limits were realized when the sample water flow-rate was maximally increased within the limits of the pump, 5–10 ml/min (despite exceeding the breakthrough threshold of the SPE microcolumn), and when the eluate volume collected from the column was minimized, <5 μl (despite very low recovery percentages).
  • Keywords
    Explosives
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2003
  • Journal title
    Journal of Chromatography A
  • Record number