• Title of article

    Analysis of polychlorinated biphenyls in aqueous samples by microwave-assisted headspace solid-phase microextraction

  • Author/Authors

    Shu، نويسنده , , Youn Yuen and Wang، نويسنده , , Shu Shing and Tardif، نويسنده , , Mylaine and Huang، نويسنده , , Yaping، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    12
  • From page
    1
  • To page
    12
  • Abstract
    The hyphenated technique namely microwave-assisted headspace solid-phase microextraction (MA-HS-SPME) was developed and studied for the simultaneous extraction/enrichment of polychlorinated biphenyls (PCBs) in aqueous samples prior to the quantification by gas chromatography (GC). The PCBs in aqueous media are extracted onto a solid-phase micro fibre via the headspace with the aid of microwave irradiation. The optimum conditions for obtaining extraction efficiency, such as the extraction time, addition of salts, addition of methanol, ratio of sample to headspace volume, and the desorption parameters were investigated. Experimental results indicated that the proposed MA-HS-SPME method attained the best extraction efficiency under the optimized conditions, i.e., irradiation of extraction solution (20 ml aqueous sample in 40 ml headspace vial with no additions of salt and methanol) under 30 W microwave power for 15 cycles (1 min power on and 3 min power off of each cycle). Desorption at 270 °C for 3 min provided the best detection results. The detection limit obtained were between 0.27 and 1.34 ng/l. The correlation coefficient for the linear dynamic range from 1 to 80 ng/l exceeded 0.99 for 18 PCBs.
  • Keywords
    Polychlorinated biphenyls
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2003
  • Journal title
    Journal of Chromatography A
  • Record number

    1517932