Title of article
Mass spectrometry-compatible ICH (International Conference on Harmonization) impurity analysis with a high-pH mobile phase: Advantages and pitfalls
Author/Authors
Dong، نويسنده , , M.W and Miller، نويسنده , , G.D and Paul، نويسنده , , R.K، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2003
Pages
8
From page
283
To page
290
Abstract
Recent advances in bonding chemistry and novel silica synthesis have significantly extended the pH range of silica-based HPLC columns. This extended range now enables the analysis of water-soluble basic drugs at high pH without ion-pairing reagents, thus offering an alternative approach to assay or impurity analyses. This paper describes the many advantages and potential pitfalls of using high-pH mobile phases in the development of MS-friendly LC gradient impurity analytical methods for water-soluble basic drugs under International Conference on Harmonization (ICH) guidelines. Operating at high-pH provides excellent peak shapes and retention, and accentuates selectivity differences of structurally similar impurities and degradants. However, several problems unique to the use of high-pH mobile phases, such as column lifetime, robustness of pH adjustments, peak fronting, and on-column dimerization, were encountered. Each of these problems is discussed with its respective remedy.