• Title of article

    Integrated moulded polymer electrodes for performing conductivity detection on isotachophoresis microdevices

  • Author/Authors

    Baldock، نويسنده , , S.J and Fielden، نويسنده , , P.R and Goddard، نويسنده , , N.J and Prest، نويسنده , , J.E and Treves Brown، نويسنده , , B.J، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2003
  • Pages
    12
  • From page
    11
  • To page
    22
  • Abstract
    The feasibility of using integrated injection moulded polymer electrodes as drive and detection electrodes for performing miniaturised isotachophoresis (ITP) separations with conductivity detection has been demonstrated. Injection moulded electrodes were produced from three different grades of carbon-filled polymer. Two of the electrode designs were found to be suitable for performing on-chip conductivity detection. The high-voltage characteristics of the microdevices were found to be suitable for performing ITP, with a power dissipation up to 1.4 W m−1 being achieved. Three model separations are presented to demonstrate the separation capability of the miniaturised injection moulded devices. Three anionic dyes, two inorganic anions and a mixture of eight alkaline earth, transition and lanthanide metal cations were analysed.
  • Keywords
    inorganic anions , Metal cations , DYES
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2003
  • Journal title
    Journal of Chromatography A
  • Record number