• Title of article

    Sample enrichment by using monolithic precolumns in microcolumn liquid chromatography

  • Author/Authors

    Lim، نويسنده , , Lee Wah and Hirose، نويسنده , , Katsuaki and Tatsumi، نويسنده , , Shinichi and Uzu، نويسنده , , Hideyuki and Mizukami، نويسنده , , Megumi and Takeuchi، نويسنده , , Toyohide، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2004
  • Pages
    8
  • From page
    205
  • To page
    212
  • Abstract
    An on-line sample enrichment system was designed using monolithic precolumns in microcolumn LC. The monolithic ODS capillary columns were prepared via in situ sol–gel processes. The enrichment efficiency of the monolithic columns was tested by using phthalates as the analytes. The relative standard deviations (n=6) for the retention time, peak area and peak height were between 0.4 and 1.2%, 0.9 and 5.5% and 0.4 and 3.9%, respectively. The system was linear (R2>0.99) within the working sample concentration and sample volume ranges. Comparing to 0.2 μl injection with a typical sample injector, the theoretical plate number of a same separation column was increased by 3–6-fold when the precolumn unit was used for sample injection. The recoveries of the analytes were between 88 and 120%, and the sample volume that could be injected into the system was increased up to 5000-fold. The limits of detection were improved by more than 2000-fold and were between 0.21 and 0.87 ng ml−1 even with a UV absorbance detector. This system was applied to the determination of phthalates contained in laboratory distilled water and tap water samples.
  • Keywords
    Sample enrichment , Water analysis , Monolithic precolumns , Phthalates
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2004
  • Journal title
    Journal of Chromatography A
  • Record number