Title of article
Determination of the isoelectric point of proteins by capillary isoelectric focusing
Author/Authors
Righetti، نويسنده , , Pier Giorgio، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2004
Pages
9
From page
491
To page
499
Abstract
Different ways of determining isoelectric points (pI) of proteins in capillary isoelectric focusing are reviewed here. Due to the impossibility of direct pH measurements in the liquid phase, such assessments have to rely on the use of pI markers. Different types of pI markers have been described: dyes, fluorescently labelled peptides, sets of proteins of known pI values. It appears that, perhaps, the best system is a set of 16 synthetic peptides, trimers to hexamers, made to contain each a Trp residue for easy detection at 280 nm. By a careful blend of acidic (Asp, Glu), mildly basic, with pK around neutrality (His), and basic (Lys, Arg) amino acids, it is possible to obtain a series of pI markers with pI values quite evenly distributed along the pH scale, possessing good buffering capacity and conductivity around their pI values and thus focusing as sharp peaks. Another approach to pI determination is the monitoring of the current during mobilization: this allows, with the aid of known pI markers, to calibrate the system with a pI/current graph. Pitfalls and common errors in pI determinations are reviewed here and guidelines given for minimizing such errors in pI estimation.
Keywords
Proteins , Isoelectric focusing , amino acids , Isoelectric points , Reviews
Journal title
Journal of Chromatography A
Serial Year
2004
Journal title
Journal of Chromatography A
Record number
1520469
Link To Document