Title of article
Low level determination of p-toluenesulfonate and benzenesulfonate esters in drug substance by high performance liquid chromatography/mass spectrometry
Author/Authors
Taylor، نويسنده , , Graham E. and Gosling، نويسنده , , Mark S Pearce، نويسنده , , Andrea، نويسنده ,
Issue Information
روزنامه با شماره پیاپی سال 2006
Pages
7
From page
231
To page
237
Abstract
GC/FID and HPLC/MS single ion monitoring methods have been evaluated for the determination of trace levels of methyl, ethyl and isopropyl esters of p-toluenesulfonic acid and methyl, ethyl, isopropyl and n-butyl esters of benzenesulfonic acid in drug substances. These sulfonate esters have been highlighted as potential genotoxins. HPLC/MS was found to be more promising and limits of quantification were between 2.5 and 5 ng/mL, which enabled detection limits in drug substance at 0.01–0.1 ppm for a 50 mg/mL solution. For one drug substance excellent recoveries of 94–95% were obtained at the 1.0 ppm level, however, with a second drug substance, a besylate salt, recoveries ranged from 86% to 100% and were dependant on the sample preparation. Limited stability of the sulfonate esters in various potential sample solutions indicated that samples may need to be prepared immediately before injection.
Keywords
Besylates , trace analysis , stability , LC HPLC/MS , HPLC , LC/MS , GC , Sulfonate esters , p-Toluenesulfonate esters , Benzenesulfonate esters , Genotoxins , Tosylates
Journal title
Journal of Chromatography A
Serial Year
2006
Journal title
Journal of Chromatography A
Record number
1521303
Link To Document