• Title of article

    Flow-through system for the generation of standard aqueous solution of polycyclic aromatic hydrocarbons

  • Author/Authors

    Ouyang، نويسنده , , Gangfeng and Chen، نويسنده , , Yong and Pawliszyn، نويسنده , , Janusz، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2006
  • Pages
    4
  • From page
    176
  • To page
    179
  • Abstract
    Standard aqueous solution of organic pollutants is used for the development of analysis methods and novel water sampling devices. The use of solid-phase microextraction (SPME) and other novel technologies for research in water sampling and analysis requires systems for the reliable generation of standard aqueous solution. In this communication, a new flow-through system based on permeation is introduced for the generation of standard aqueous solution of polycyclic aromatic hydrocarbons (PAHs). The main difference between the new developed system and typical system is that the syringe pump was changed with a permeation generator. The generator-based permeation offers many advantages, including convenient, inexpensive, solvent-free, long lifetime and production of high concentrations of the target analytes. The observed change in the concentration of the six PAH compounds was less than 20% over 3 months.
  • Keywords
    flow-through system , permeation , passive sampling , Polycyclic Aromatic Hydrocarbons (PAHs)
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2006
  • Journal title
    Journal of Chromatography A
  • Record number