• Title of article

    Analysis of linear and cyclic oligomers in polyamide-6 without sample preparation by liquid chromatography using the sandwich injection method: I. Injection procedure and column stability

  • Author/Authors

    Mengerink، نويسنده , , Y and Peters، نويسنده , , R and Kerkhoff، نويسنده , , M and Hellenbrand، نويسنده , , Jikke M.T. Omloo، نويسنده , , H and Andrien، نويسنده , , J and Vestjens، نويسنده , , M and van der Wal، نويسنده , , Sj، نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2000
  • Pages
    14
  • From page
    37
  • To page
    50
  • Abstract
    We report a method for reliable routine polymer sample introduction with minimal bias, a separation method of the first six linear and cyclic oligomers by liquid chromatography, quantification using group equivalents and long term method performance. Injecting a polymer sample in a mobile phase containing an aqueous non-solvent often results in blocked systems as the polymer precipitates in the connecting capillaries. In this first part we focus on a new injection technique, in which the dissolved polyamide is placed between two zones of formic acid, preventing the polymer to precipitate before it reaches the column. Development of this sandwich injection method makes direct injection of the polymer into an aqueous acetonitrile gradient feasible. The oligomeric polyamide recovery of this technique, extraction, dissolution/precipitation and direct injection on a hexafluoro–isopropanol (HFIP) gradient were compared. With the sandwich injection method the polymer remains on the column, slowly changing the stationary phase. The influence of this on resolution and retention was studied. Column stability allows sixty injections before cleaning or replacing the column is necessary.
  • Keywords
    Nylon , polyamide , Polymer
  • Journal title
    Journal of Chromatography A
  • Serial Year
    2000
  • Journal title
    Journal of Chromatography A
  • Record number

    1527240