    Determination of the presampled MTF in computed tomography

  • Author/Authors

    Boone، John M. نويسنده ,

  • Issue Information
    روزنامه با شماره پیاپی سال 2001
  • Pages
    -355
  • From page
    356
  • To page
    0
  • Abstract
    A technique for measuring the presampled MTF in CT scanners is described. The technique uses a simple phantom consisting of ~0.050 mm aluminum foil sandwiched by flat plastic or tissue-equivalent slabs. The aluminum foil is slightly angled with respect to the reconstruction matrix, and CT images are acquired. The acquired CT image yields an angled slit image that can be used to synthesize the presampled line spread function (LSF). The presampled MTF is calculated from the presampled LSF. The technique is a direct extension of that proposed by Fujita et al. [IEEE Trans. Med. Imaging 11, 34-39 (1992)] for MTF calculation on digital radiography images. While the MTF in clinical CT scanners often reaches negligible amplitude below the Nyquist frequency, the technique is easy to implement, requires inexpensive materials, is robust to aliasing, and is more resilient to noise due to greater data averaging than conventional PSF-integration techniques. Use of the proposed technique is illustrated on a clinical multiple detector array scanner, and MTFs are shown for several common reconstruction kernels. It is likely that the proposed technique would be useful for all tomographic imaging systems, including single photon emission computed tomography, positron emission tomography, magnetic resonance imaging and ultrasound scanners.
  • Keywords
    Fault current limiter , short circuit current , transient over voltage , power quality
  • Journal title
    MEDICAL PHYSICS
  • Serial Year
    2001
  • Journal title
    MEDICAL PHYSICS
  • Record number

    1529