• Title of article

    Determination of water traces in various organic solvents using Karl Fischer method under FIA conditions

  • Author/Authors

    Dantan، نويسنده , , Nathalie and Frenzel، نويسنده , , Wolfgang and Küppers، نويسنده , , Stephan، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2000
  • Pages
    9
  • From page
    101
  • To page
    109
  • Abstract
    Flow injection methods utilising the Karl Fischer (KF) reaction with spectrophotometric and potentiometric detection are described for the determination of the trace water content in various organic solvents. Optimisation of the methods resulted in an accessible (linear) working range of 0.01–0.2% water for many solvents studied with a typical precision of 1–2% R.S.D. Only 50 μl of organic solvent was injected and the sampling frequency was about 120 samples per h. Since the slopes of the calibration curves were different for different solvents appropriate calibration was required. Problems associated with spectrophotometric detection and caused by refractive index changes were pointed out and a nested-loop configuration was proposed to overcome this kind of interference. The potentiometric method with a novel flow-through detector cell was shown to surpass the performance of spectrophotometric detection in any respect. The characteristics of the procedures developed made them well applicable for on-line monitoring of technical solvent distillations in an industrial plant.
  • Keywords
    Flow injection analysis , Karl Fischer method , Spectrophotometry , Potentiometry , Process control , Organic solvents , Water determination
  • Journal title
    Talanta
  • Serial Year
    2000
  • Journal title
    Talanta
  • Record number

    1640257