• Title of article

    Simultaneous spectrophotometric determination of diuretics by using multivariate calibration methods

  • Author/Authors

    Luis، نويسنده , , M.L and Fraga، نويسنده , , J.M.G and Jiménez، نويسنده , , F and Jiménez، نويسنده , , A.I and Arias، نويسنده , , J.J، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2001
  • Pages
    10
  • From page
    761
  • To page
    770
  • Abstract
    The wavelength range and number of factors used in partial least-squares (PLS) calibration for the resolution of the dihydralazine (DHZ)–hydrochlorothiazide (HCT) binary mixture and the dihydralazine–hydrochlorothiazide–reserpine ternary mixture were optimized in terms of the relative standard error (R.S.E.) and relative mean standard error (R.M.S.E.). Under the optimum conditions thus established, synthetic mixtures of the analytes can be resolved with errors and relative standard deviations (R.S.D) less than 4.5 and 1.0%, respectively. The ensuing method, which was validated by comparison with high performance liquid chromatography (HPLC), also gives good results with real samples (pharmaceutical preparations).
  • Keywords
    Wavelength range selection , Partial least-squares , Diuretics
  • Journal title
    Talanta
  • Serial Year
    2001
  • Journal title
    Talanta
  • Record number