• Title of article

    A novel tetrachlorothallate (III)–PVC membrane sensor for the potentiometric determination of thallium (III)

  • Author/Authors

    Hassanien، نويسنده , , M.M and Abou-El-Sherbini، نويسنده , , Kh.S and Mostafa، نويسنده , , G.A.E، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2003
  • Pages
    10
  • From page
    383
  • To page
    392
  • Abstract
    A novel tetrachlorothallate (III) (TCT)-selective membrane sensor consisting of tetrachlorothallate (III)–2,3,5-triphenyl-2-H-tetrazolium ion pair dispersed in a PVC matrix plasticized with dioctylphthalate is described. The electrode shows a stable, near-Nernstian response for 1×10−3–4×10−6 M thallium (III) at 25 °C with an anionic slope of 56.5±0.5 over the pH range 3–6. The lower detection limit and the response time are 2×10−6 M and 30–60 s, respectively. Selectivity coefficients for Tl(III) relative to a number of interfering substances were investigated. There is negligible interference from many cations and anions; however, iodide and bromide are significantly interfere. The determination of 0.5–200 μg ml−1 of Tl(III) in aqueous solutions shows an average recovery of 99.0% and a mean relative standard deviation of 1.4% at 50.0 μg ml−1. The direct determination of Tl(III) in spiked wastewater gave results that compare favorably with those obtained by the atomic absorption spectrometric method. The electrode was successfully applied for the determination of thallium in zinc concentrate. Also the tetrachlorothallate electrode has been utilized as an end point indicator electrode for the determination of thallium using potentiometric titration.
  • Keywords
    2 , 3 , 5-Triphenyl-2-H-tetrazolium , PVC Membrane Electrode , Tetrachlorothalate (III)
  • Journal title
    Talanta
  • Serial Year
    2003
  • Journal title
    Talanta
  • Record number

    1643756