• Title of article

    Determination of total chromium traces in tannery effluents by electrothermal atomic absorption spectrometry, flame atomic absorption spectrometry and UV–visible spectrophotometric methods

  • Author/Authors

    Monteiro، نويسنده , , M.I.C and Fraga، نويسنده , , I.C.S and Yallouz، نويسنده , , A.V and de Oliveira، نويسنده , , N.M.M. and RIBEIRO، نويسنده , , S.H، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2002
  • Pages
    5
  • From page
    629
  • To page
    633
  • Abstract
    Three different analytical methods comprising colorimetric method with 1,5-diphenyl-carbazide, electrothermal atomic absorption spectrometry (ET AAS) and flame atomic absorption spectrometry were utilized in a study to determine traces of chromium (Cr) in synthetic tannery effluent from laboratory scale treatment process variations. All the results obtained using the three different methods showed good agreement and met the requirement of Brazilian regulation for total Cr for effluent discharges (<0.5 mg l−1). However, ET AAS has been the proposed method because it was faster, less laborious, needed smaller volume of sample and presented lower limit of quantification (LOQ=2.2 μg l−1).
  • Keywords
    Colorimetric method , Electrothermal atomic absorption spectrometry , Total chromium , Flame atomic absorption spectrometry , Tannery effluents
  • Journal title
    Talanta
  • Serial Year
    2002
  • Journal title
    Talanta
  • Record number