• Title of article

    Flow injection system with in-line Winkler’s procedure using 16-way valve and spectrofluorimetric determination of dissolved oxygen in environmental waters

  • Author/Authors

    Sakai، نويسنده , , Tadao and Piao، نويسنده , , Shanji and Teshima، نويسنده , , Norio and Kuroishi، نويسنده , , Tadafumi and Grudpan، نويسنده , , Kate، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    893
  • To page
    898
  • Abstract
    Flow injection spectrofluorimetry with in-line Winklers procedure was developed for the dissolved oxygen (DO) determination. 2-Thionaphthol reacted with iodine produced by Winkler’s method to form fluorescence inactive disulfide compound. To automate the process completely, a 5-channel flow system with a newly designed 16-way valve was assembled. The system consisted of a dispersion coil (DC), a precipitate formation coil (PFC), a precipitate dissolving coil (PDC), and extraction coil (EC). A calibration can be constructed by using a standard iodine solution for dissolved oxygen. The calibration graph was linear over the range 1.2×10−4∼6.0×10−4 mol l−1 iodine (1.96∼9.80 mg O l−1)). The relative standard deviation (n=6) was below 0.3% for the 4×10−4 mol l−1 iodine (6.27 mg O l−1) determination. The sample throughput was 12/h.
  • Keywords
    Dissolved oxygen , 2-Thionaphthol , In-line reaction , Flow injection analysis , Winkler’s method , Fluorimetry , 16-Way valve
  • Journal title
    Talanta
  • Serial Year
    2004
  • Journal title
    Talanta
  • Record number

    1646005