• Title of article

    Strategies for constructing the calibration set for a near infrared spectroscopic quantitation method

  • Author/Authors

    Blanco، نويسنده , , M. and Romero، نويسنده , , M.A. and Alcalà، نويسنده , , M.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2004
  • Pages
    6
  • From page
    597
  • To page
    602
  • Abstract
    Three strategies for the construction of calibration sets have been tried, with the objective to develop and to validate a NIR quantitation method. rst two approaches consist of the use of two types of samples, named: samples of laboratory obtained by mixing the ingredients that compose the drug, and doped samples obtained by under- and over-dosed production samples. In order to improve the prediction results, production samples have been added to each calibration model. The ensuing models were validated with a view to determine their fitness for purpose. However, spectral differences between the laboratory samples and doped samples resulted in spurious predictions in quantifying samples of one type using the model developed from samples of the other. ifferences were studied in depth and a third procedure has been proposed, based on a calibration model constructed with an unique type of sample (laboratory sample) for later to correct it with a few doped samples. This corrected model has a good predictive ability on production samples.
  • Keywords
    near infrared spectroscopy , Calibration , NIR , Pharmaceutical Analysis
  • Journal title
    Talanta
  • Serial Year
    2004
  • Journal title
    Talanta
  • Record number