• Title of article

    Separation of PAHs in aerosol by thin layer chromatography for compound-specific stable carbon isotope analysis

  • Author/Authors

    Liu، نويسنده , , Xiaohong and Bi، نويسنده , , Xinhui and Mai، نويسنده , , Bixian and Sheng، نويسنده , , Guoying and Fu، نويسنده , , Jiamo، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2005
  • Pages
    8
  • From page
    487
  • To page
    494
  • Abstract
    A method using a thin layer chromatography (TLC) for compound-specific stable carbon isotope analysis of polycyclic aromatic hydrocarbons (PAHs) with four to seven rings was developed in this study. Five aerosol samples were used as test samples. Two stationary phases and eight developing systems were tested. The results indicated: (1) silica gel is superior to aluminum oxide and the silica gel precoated plate developed with hexane:chloroform (45:5, v/v) can give the best separation effect; (2) individual PAHs associated with aerosols can be effectively separated from unresolved complex mixture (UCM) by this procedure. The carbon isotope composition of PAHs can be measured with a standard deviation (S.D.) < 0.5‰, n = 4. No significant isotopic fractionation was observed during the TLC procedure. And this technique can be used as a potential tool for source identification of PAHs in the aerosols.
  • Keywords
    aerosol , thin layer chromatography , Carbon isotope , Separation , PAHS
  • Journal title
    Talanta
  • Serial Year
    2005
  • Journal title
    Talanta
  • Record number

    1648057