MCM-41 mesoporous material modified carbon paste electrode for the determination of cardiac troponin I by anodic stripping voltammetry
Author/Authors
Guo، نويسنده , , Huishi and He، نويسنده , , Nongyue and Ge، نويسنده , , Shuxun and Yang، نويسنده , , Di and Zhang، نويسنده , , Jinan، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2005
Pages
6
From page
61
To page
66
Abstract
An anodic stripping voltammetric method for the determination of cardiac troponin I (cTnI) at a MCM-41 mesoporous material modified carbon paste electrode (MCM-MCPE) was investigated. The test was based on the dual monoclonal antibody “sandwich” principle using colloidal gold as a labeled substrate. Four main steps were carried out to obtain the analytical signal, i.e. electrode preparation, immunoreaction, silver enhancement, and anodic stripping voltammetric detection. The anodic stripping peak current increased linearly with the concentration of cTnI over the range of 0.8–5.0 ng/ml. A detection limit of 0.5 ng/ml was obtained. The established method was applied to detect cTnI in acute myocardial infarction (AMI) samples using routine enzyme-linked immunoadsorbent assay (ELISA) for comparison analysis, and good results were obtained.
Keywords
Cardiac troponin I , mesoporous materials , MCM-41 , Anodic stripping voltammetry , Carbon paste electrode