• Title of article

    Uranium colloid analysis by single particle inductively coupled plasma-mass spectrometry

  • Author/Authors

    Degueldre، نويسنده , , C. and Favarger، نويسنده , , P.-Y. and Rossé، نويسنده , , R. and Wold، نويسنده , , S.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2006
  • Pages
    6
  • From page
    623
  • To page
    628
  • Abstract
    Uranium single particle analysis has been performed by inductively coupled plasma-mass spectrometry (ICP-MS) and the performances are compared with that provided by scanning electron microsopy and single particle counting. The transient signal induced by the flash of ions due to the ionisation of an uranium colloidal particle in the plasma torch can be detected and measured for selected uranium ion masses (238U+, 235U+ or 254[238U16O]+) by the mass spectrometer. The signals recorded via time scanning are analysed as a function of particle size or fraction of the studied element or isotope in the colloid phase. The frequency of the flashes is directly proportional to the concentration of particles in the colloidal suspension. The feasibility tests were performed on uranium dioxide particles. The study also describes the experimental conditions and the choice of mass to detect uranium colloids in a single particle analysis mode.
  • Keywords
    Particle size , isotope , Uranium particle , INDUCTIVELY COUPLED PLASMA-MASS SPECTROMETRY , Single particle analysis
  • Journal title
    Talanta
  • Serial Year
    2006
  • Journal title
    Talanta
  • Record number