• Title of article

    Validation of mercury determination by solid sampling Zeeman atomic absorption spectrometry and a specially designed furnace

  • Author/Authors

    Karl Heinz Grobecker، نويسنده , , Karl-Heinz and Detcheva، نويسنده , , Albena، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2006
  • Pages
    4
  • From page
    962
  • To page
    965
  • Abstract
    Certified reference materials (CRMs) of different origin were used to validate the direct determination of total mercury by solid sampling Zeeman atomic absorption spectrometry (SS-ZAAS) and a specially designed furnace. The temperature program provides only for one step. Atomisation of mercury and pyrolysis of the matrix is performed at a constant temperature in the range of 900–1000 °C. Calibration points achieved by CRMs and aqueous solutions are covered by one calibration line, indicating the absence of matrix effects. Relatively high amounts of chlorine, known for causing problems in mercury determination do not influence analytical results. The excellent accuracy of the method results in a very good agreement with the certified values. The precision of SS-ZAAS measurements in a range from 0.5 to 50 ng Hg does not exceed 3% R.S.D. A limit of quantification of 0.008 μg g−1 Hg was achieved.
  • Keywords
    Method validation , mercury , Matrix interferences , Specially designed furnace , Solid sampling Zeeman atomic absorption spectrometry
  • Journal title
    Talanta
  • Serial Year
    2006
  • Journal title
    Talanta
  • Record number