• Title of article

    Flow-injection chemiluminescence determination of formaldehyde in water

  • Author/Authors

    Motyka، نويسنده , , Kamil and Onjia، نويسنده , , Antonije and Miku?ka، نويسنده , , Pavel and Ve?e?a، نويسنده , , Zbyn?k، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2007
  • Pages
    6
  • From page
    900
  • To page
    905
  • Abstract
    A modification of the Trautz–Schorigin reaction into a flow-injection analysis configuration is described. Different approaches were used at the optimization of chemiluminescence determination of formaldehyde in water based on the reaction of formaldehyde, gallic acid and hydrogen peroxide in an alkaline solution. Detection system with a 218 μl chemiluminescence cell was optimized by both a one-variable-at-a-time method, and a modified simplex method. A calibration graph is linear in the concentration range 4 × 10−8 to 1 × 10−5 M HCHO. The detection limit of formaldehyde for a signal-to-noise ratio of 3 is 4 × 10−8 M. The relative standard deviations for 15 repeated measurements of 1 × 10−6 and 5 × 10−6 mol l−1 HCHO are 4.32 and 3.33%, respectively. The analysis time is 1.5 min. The method was applied to the determination of formaldehyde in urban rainwater. A comparison of results found by proposed method with those obtained by fluorimetric reference method provided a good agreement.
  • Keywords
    Formaldehyde , Chemiluminescence , Flow-injection analysis , Simplex
  • Journal title
    Talanta
  • Serial Year
    2007
  • Journal title
    Talanta
  • Record number