Title of article
An ion-exchange method for speciation of antimony by flow injection electrothermal atomic absorption spectrometry
Author/Authors
Sلnchez Rojas، نويسنده , , F. and Bosch Ojeda، نويسنده , , C. and Cano Pavَn، نويسنده , , J.M.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2007
Pages
5
From page
918
To page
922
Abstract
In this work, a simple preconcentration system, achieved by replacing the sample tip of the autosampler arm by a micro-column packed with Amberlite IRA-910 or silica gel chelating resin functionalised with 1,5-bis(di-2-pyridyl)methylene tbiocarbohydrazide (DPTH-gel), is developed for the determination of Sb(V) and total antimony, respectively. Different factors including pH of sample solution, ionic strength, concentration and volume of eluent, sample flow rate, sample loading time and matrix effects for preconcentration were investigated. The method has been applied to the determination of antimony species in different samples.
Keywords
Electrothermal atomic absorption spectrometry , Flow Injection , Speciation , Antimony determination , Preconcentration
Journal title
Talanta
Serial Year
2007
Journal title
Talanta
Record number
1651393
Link To Document