• Title of article

    Optimisation of ICPMS collision/reaction cell conditions for the simultaneous removal of argon based interferences of arsenic and selenium in water samples

  • Author/Authors

    Darrouzès، نويسنده , , Jérôme and Bueno، نويسنده , , Maïté and Lespès، نويسنده , , Gaëtane and Holeman، نويسنده , , Michel and Potin-Gautier، نويسنده , , Martine، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2007
  • Pages
    5
  • From page
    2080
  • To page
    2084
  • Abstract
    The optimisation of ICPMS collision/reaction cell conditions for the simultaneous analysis of arsenic and selenium is described. A mixture of 3.8 mL min−1 of H2 and 0.5 mL min−1 of He was found to be suitable for the removal of both ArAr+ and ArCl+ interferences. Detection limits down to 30 ng (element) L−1 in total analysis, and between 81 and 230 ng (element) L−1 in speciation analysis were achieved in chloride matrix (1 g L−1 NaCl). After validation, the method was applied to commercially available mineral waters.
  • Keywords
    Selenium , Speciation , ICPMS , Reaction cell , water , Arsenic , Collision cell , Octopole
  • Journal title
    Talanta
  • Serial Year
    2007
  • Journal title
    Talanta
  • Record number