Title of article
Greener analytical method for the determination of copper(II) in wastewater by micro flow system with optical sensor
Author/Authors
Leelasattarathkul، نويسنده , , Tapparath and Liawruangrath، نويسنده , , Saisunee and Rayanakorn، نويسنده , , Mongkon and Liawruangrath، نويسنده , , Boonsom and Oungpipat، نويسنده , , Winai and Youngvises، نويسنده , , Napaporn، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2007
Pages
6
From page
126
To page
131
Abstract
Greener analytical method using micro flow system for the determination of Cu(II) in wastewater samples was designed and investigated. The micro flow system consisted of a planar glass chip with poly(dimethylsiloxane) (PDMS) top plate and fixed with fiber optic probe as optical sensor for monitoring of Cu(II) that reacted with 2-carboxy-2′-hydroxy-5′-sulfoformazyl benzene (zincon) on the chip at 605 nm. This design gave a satisfied sensitivity with a linear calibration graph over the range of 0.1–3.0 μg mL−1 of Cu(II) and correlation coefficient 0.9991. The percentage relative standard deviation was 2.5 for 10-replicate measurements and the limit of detection (LOD) was 0.1 μg mL−1. This system has been successfully applied to the determination of Cu(II) in wastewaters from electroplating industry with less reagents and samples consumption and diminutive waste generation.
Keywords
Green analytical chemistry , Optical sensor , Copper(II) , Micro flow injection