Title of article
A new hydride generation system applied in determination of arsenic species with ion chromatography–hydride generation-atomic fluorescence spectrometry (IC–HG-AFS)
Author/Authors
Wei، نويسنده , , Changjin and Liu، نويسنده , , Jixin، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2007
Pages
6
From page
540
To page
545
Abstract
A novel procedure was developed for the determination of arsenite (As(III)), arsenate (As(V)), monomethylarsonic (MMA) and dimethylarsinic acid (DMA) with ion chromatography–hydride generation-atomic fluorescence spectrometry (IC–HG-AFS) by employing a new gas–liquid separator (GLS). The effective separation of the four arsenic species was achieved in about 12 min. With a sample loading volume of 20 μl, the measurable minimum for As(III), DMA, MMA and As(V) were 0.02, 0.045, 0.043 and 0.166 ng, respectively, along with relative standard deviations of 1.1, 1.1, 1.7 and 2.2% at the 100 μg l−1 level (n = 6) for As(III), DMA, MMA and As(V), respectively. The present procedure was applied for the speciation of arsenic in underground water and in urine samples, and the sum of the four arsenic species by IC–HG-AFS was in good agreement with the total value by HG-AFS.
Keywords
Ion chromatography , Hydride generation-atomic fluorescence spectrometry , Gas–liquid separator , Arsenic speciation
Journal title
Talanta
Serial Year
2007
Journal title
Talanta
Record number
1653037
Link To Document