• Title of article

    Enzymatic probe sonication as a tool for solid–liquid extraction for total selenium determination by electrothermal-atomic absorption spectrometry

  • Author/Authors

    Vale، نويسنده , , G. and Pereira، نويسنده , , Claudenilson S. and Mota، نويسنده , , A. and Fonseca، نويسنده , , L. and Capelo، نويسنده , , J.L.، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2007
  • Pages
    8
  • From page
    198
  • To page
    205
  • Abstract
    A new fast and reproducible approach is described for the application of the enzymatic probe sonication (EPS) methodology [J.L. Capelo, P. Ximénez-Embún, Y. Madrid-Albarrán, C. Cámara, Anal. Chem. 76 (2004) 233–237] for total selenium determination by electrothermal atomic absorption spectrometry, ET-AAS. Ni(NO3)2 and Pd(NO3)2 were studied as matrix modifiers in conjunction with H2O2, being best results obtained with Pd(NO3)2 plus H2O2. The presence of H2O2 as matrix modifier increases up to 66% the time-life of the graphite tubes, by avoiding the building-up of carbonaceous residues. BCR-414 plankton and ERM-CE 278 mussel tissue reference materials were used for proof-of-the-methodology. Different enzymes, protease XIV, substilisin and trypsin were studied. The use of fresh enzyme was found critical. Good Se recoveries were obtained for oyster tissue, 111%; BCR-414 plankton, 106% and ERM-CE 278 mussel tissue, 93%, when protease XIV was used. Data regarding microwave digestion versus EPS methodology is also presented and discussed.
  • Keywords
    Solid–liquid extraction , Selenium , EPS , ET-AAS , enzymes , focused ultrasound
  • Journal title
    Talanta
  • Serial Year
    2007
  • Journal title
    Talanta
  • Record number

    1653321