Title of article
Graphite–silicone rubber composite electrode: Preparation and possibilities of analytical application
Author/Authors
de Oliveira، نويسنده , , Aline Carlos and dos Santos، نويسنده , , Sidney Xavier and Cavalheiro، نويسنده , , ةder Tadeu Gomes، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2008
Pages
7
From page
1043
To page
1049
Abstract
Composite electrodes were prepared using graphite powder and silicone rubber in different compositions. The use of such hydrophopic materials interned to diminish the swallowing observed in other cases when the electrodes are used in aqueous solutions for a long time. The composite was characterized for the response reproducibility, ohmic resistance, thermal behavior and active area. The voltammetric response in relation to analytes with known voltammetric behavior was also evaluated, always in comparison with the glassy carbon. The 70% (graphite, w/w) composite electrode was used in the quantitative determination of hydroquinone (HQ) in a DPV procedure in which a detection limit of 5.1 × 10−8 mol L−1 was observed. HQ was determined in a photographic developer sample with errors lower then 1% in relation to the label value.
Keywords
Composite electrode material , Ferricyanide , Graphite , hydroquinone , silicone rubber
Journal title
Talanta
Serial Year
2008
Journal title
Talanta
Record number
1653980
Link To Document