• Title of article

    An aqueous two-phase system as a strategy for serum albumin depletion

  • Author/Authors

    da Silva، نويسنده , , Marcelo Anselmo Oseas and Arruda، نويسنده , , Marco Aurélio Zezzi Arruda، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2009
  • Pages
    6
  • From page
    985
  • To page
    990
  • Abstract
    An aqueous two-phase system, based on the cloud point phenomenon, was proposed for blood plasma albumin depletion prior to gel electrophoresis (SDS-PAGE) analysis. Albumin removal was evaluated using a mixture of Triton® X-114 and SDS. Variables affecting phase separation and the partition coefficient (K), such as pH (from 5.0 to 7.6) and salt composition, were first optimized to carry out phase separation at room temperature and to promote albumin removal (15 min with ca. 95% efficiency) to the surfactant-rich phase, in a global process that results in reduction of 12% α-helix structures, confirmed through circular dichroism analysis. Gel electrophoresis analyses were carried out to compare crude human plasma composition with human plasma submitted to the cloud point procedure as well as the same plasma submitted to the separation of albumin using a commercially available system based on a chromatographic affinity column. The cloud point extraction procedure allowed identification of 18 protein bands present in the surfactant-poor phase, resulting in a more detailed electrophoretic profile, when compared with crude plasma (10 protein bands).
  • Keywords
    Albumin depletion , circular dichroism , Gel Electrophoresis , Cloud point
  • Journal title
    Talanta
  • Serial Year
    2009
  • Journal title
    Talanta
  • Record number

    1656242