• Title of article

    Dispersive liquid–liquid microextraction preconcentration of palladium in water samples and determination by graphite furnace atomic absorption spectrometry

  • Author/Authors

    Liang، نويسنده , , Pei and Zhao، نويسنده , , Ehong and Li، نويسنده , , Feng، نويسنده ,

  • Issue Information
    ماهنامه با شماره پیاپی سال 2009
  • Pages
    4
  • From page
    1854
  • To page
    1857
  • Abstract
    A new method for the determination of palladium was developed by dispersive liquid–liquid microextraction preconcentration and graphite furnace atomic absorption spectrometry detection. In the proposed approach, diethyldithiocarbamate (DDTC) was used as a chelating agent, and carbon tetrachloride and ethanol were selected as extraction and dispersive solvent. Some factors influencing the extraction efficiency of palladium and its subsequent determination, including extraction and dispersive solvent type and volume, pH of sample solution, concentration of the chelating agent and extraction time, were studied and optimized. Under the optimum conditions, the enrichment factor of this method for palladium reached at 156. The detection limit for palladium was 2.4 ng L−1 (3σ), and the relative standard deviation (R.S.D.) was 4.3% (n = 7, c = 1.0 ng mL−1). The method was successfully applied to the determination of trace amount of palladium in water samples.
  • Keywords
    PALLADIUM , Dispersive liquid–liquid microextraction , Graphite furnace atomic absorption spectrometry , Water samples