A simple electrochemical method was developed for the single and simultaneous determination of butylated hydroxyanisole (BHA) and butylated hydroxytoluene (BHT) in food samples using square-wave voltammetry (SWV). A carbon composite electrode modified (MCCE) with copper (II) phosphate immobilized in a polyester resin was proposed. The modified electrode allowed the detection of BHA and BHT at potentials lower than those observed at unmodified electrodes. A separation of about 430 mV between the peak oxidation potentials of BHA and BHT in binary mixtures was obtained. The calibration curves for the simultaneous determination of BHA and BHT demonstrated an excellent linear response in the range from 3.4 × 10−7 to 4.1 × 10−5 mol L−1 for both compounds. The detection limits for the simultaneous determination of BHA and BHT were 7.2 × 10−8 and 9.3 × 10−8 mol L−1, respectively. In addition, the stability and repeatability of the electrode were determined. The proposed method was successfully applied in the simultaneous determination of BHA and BHT in several food samples, and the results obtained were found to be similar to those obtained using the high performance liquid chromatography method with agreement at 95% confidence level.
Keywords
Modified carbon composite electrode , Butylated hydroxyanisole (BHA) , Butylated hydroxytoluene (BHT) , Copper (II) phosphate immobilized
Journal title
Talanta
Serial Year
2010
Journal title
Talanta
Record number
1659960
Link To Document