Title of article
Spectrophotometric determination of zinc and copper in a multi-syringe flow injection analysis system using a liquid waveguide capillary cell: Application to natural waters
Author/Authors
Pلscoa، نويسنده , , Ricardo N.M.J. and Tَth، نويسنده , , Ildikَ V. and Rangel، نويسنده , , Antَnio O.S.S.، نويسنده ,
Issue Information
ماهنامه با شماره پیاپی سال 2011
Pages
6
From page
1267
To page
1272
Abstract
This work exploits a multi-syringe injection analysis (MSFIA) system coupled with a long liquid waveguide capillary cell for the spectrophotometric determination of zinc and copper in waters. A liquid waveguide capillary cell (1.0 m pathlength, 550 μm i.d. and 250 μL internal volume) was used to enhance the sensitivity of the detection. The determination for both ions is based on a colorimetric reaction with zincon at different pH values. The developed methodology compares favourably with other previously described procedures, as it allows to reach low detection limits for both cations (LODs of 0.1 and 2 μg L−1, for copper and zinc, respectively), without the need for any pre-concentration step. The system also provided a linear response up to 100 μg L−1 with a high throughput (43 h−1) and low reagent consumption and effluent production. The developed work was applied to natural waters and three certified reference water samples.
Keywords
Spectrophotometry , Zinc , Liquid waveguide capillary cell , Multi-syringe flow injection analysis , Copper , natural waters
Journal title
Talanta
Serial Year
2011
Journal title
Talanta
Record number
1662408
Link To Document